Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)
BioFire Diagnostics
Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
The plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.
CoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US;
Medtronic Heart Valves Division
Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020.
Atellica IM 1600 Analyzer- multi-component system for in vitro diagnostic testing of clinical specimens SMN 11066000
Siemens Healthcare Diagnostics
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
VISIONIST CRT-P pacemaker, Models: a) U225 b) U226 c) U228
Boston Scientific
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.
ACE Control Set - IVD Control set for the quantitative determination of Angiotensin Converting Enzyme (ACE) REF 1667001
Sentinel CH SpA
Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.
Atellica CH 930 Analyzer - multi-component system for in vitro diagnostic testing of clinical specimens SMN 11067000
Siemens Healthcare Diagnostics
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (30 test kit) Part No: RFIT-ASY-0142 / UDI: 00815381020321 (6 test kit)
BioFire Diagnostics
Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
Potential for a false-negative result.
When adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.
Proponent pacemaker, Models: a) L200 b) L201 c) L209 d) L210 e) L211 f) L221 g) L231
Boston Scientific
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728332 - IQon Spectral CT : IQon Spectral CT
Philips North America
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882446 - Vereos PET/CT
Philips North America
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098
Magellan Diagnostics
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
There is a potential that the pressure tubing may detach from the blood sampling system
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/728201
Philips North America
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Synaptive Trackable Suction Set Standard and Malleable
Synaptive Medical
Due to visible burrs/metal filings affixed along the suction tube inner perimeter.
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Philips North America
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
Edwards Lifesciences
There is a potential that the pressure tubing may detach from the blood sampling system