"Chattering" when cutting burrs are used in bone.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
HomeSafe AutoAlert Pendant works in conjunction with a compatible Lifeline communicator.
Lifeline Systems Company
A programing error in some Model FD100 HomeSafe AutoAlert Pendants will render the fall detection feature inoperable.
Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard Technology, Part Numbers: (1) ASK-45541-RWJ4; (2) ASK-45552-RWJ4
Arrow International
Certain lots of cardiovascular catheters may not be completely sealed and so sterility of the product cannot be guaranteed.
Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.
Fresenius Medical Care Renal Therapies Group
The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.
LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Magellan Diagnostics
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Dynavox T10 Speech Generating Device ( including Batteries)
Dynavox Systems
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Seiler Instr. & Mfg. Co.
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
A.L.P.S. MIS Calcaneus Locking Plate LG RT F.A.S.T. Guide inserts, Item Number: 816209003,UDI: 00887868044670989403
Zimmer Biomet
The drill guides were manufactured with the incorrect color anodization.
DS Titanium Ligation Clip, used in laparoscopic surgery
Aesculap Implant Systems
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm.
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
Pneumothorax Kit
Arrow International
Product packaging may not be completely sealed.
Dynavox T15 Speech Generating Device ( including Batteries)
Dynavox Systems
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
Large Volume Abdominal Paracentesis Kit
Arrow International
Product packaging may not be completely sealed.
2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) ndicated for acute and chronic dialysis therapy.
Fresenius Medical Care Renal Therapies Group
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.
Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers and could result in visually underestimating calculated Ktrans, Kep, and iAUGC values. The defect also impacts DynaCAD s on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps. If the ADC values are computed on-the-fly by DynaCAD, the ADC values and colors will also display incorrectly if viewed on a remote DynaCAD client.
ArrowClarke" PleuraSeal¿ Thoracentesis Kit
Arrow International
Product packaging may not be completely sealed.
DS Titanium Ligation Clip Appliers, used in laparoscopic surgery
Aesculap Implant Systems
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. Updated cleaning and sterilization instructions.
The AIM Tibial Nails failed a clinical analysis where the surgical revision rate exceeded the acceptance criteria in the Device Performance Review protocol for the implant system.
COACH¿ Self-Adhering Sports Wrap (size: 2in)
Johnson & Johnson Consumer
Johnson & Johnson Consumer Inc. is initiating a voluntary, Class II recall at the retail level in the U.S. of select COACH¿ Self-Adhering Sports Wraps, BAND-AID¿ Brand First Aid Products SECURE-FLEX¿ Wraps and BAND-AID¿ Brand First Aid Products HURT-FREE¿ Wraps products. The current labeling claim states not made with natural rubber latex which needs to be updated based upon recent awareness that natural latex was used as a base ingredient in the early-stage manufacturing process, which reduces the allergic protein found in natural latex.
IMRIS ORT300, Removable Operating Room Table, Part Numbers: (a) 114093-000 (b) 114093-600
Deerfield Imaging
The Operating Room Table may drift while in use, which could potentially result in unintended movement of the table.